CVE-2022-4992

8.6 HIGH
Published: June 02, 2026 Modified: June 04, 2026
View on NVD

Description

DrΓ€ger Infinity Acute Care System and Standalone Infinity M540 patient monitors versions VG4.1.1, VG4.0.3, and lower (with VG4.2 partially affected) contain a network message handling vulnerability that allows remote attackers to inject spoofed or tampered data and cause denial-of-service conditions. Attackers can compromise network communications to modify device settings such as alarm states or alarm limits, or overwhelm the system with excessive network traffic causing the Cockpit or M540 to reboot and lose network functionality.

AI Explanation

Get an AI-powered plain-language explanation of this vulnerability and remediation steps.

Login to generate AI explanation

CVSS v3.x Details

0.0 Low Medium High Critical 10.0
Vector String
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H

References to Advisories, Solutions, and Tools

Patch Vendor Advisory Exploit Third Party Advisory

2 reference(s) from NVD

Quick Stats

CVSS v3 Score
8.6 / 10.0
EPSS (Exploit Probability)
0.1%
19th percentile
Exploitation Status
Not in CISA KEV

Weaknesses (CWE)