CVE-2026-7251

9.8 CRITICAL
Published: May 26, 2026 Modified: June 04, 2026
View on NVD

Description

Eppendorf BioFlo 320Β is vulnerable due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.

AI Explanation

Get an AI-powered plain-language explanation of this vulnerability and remediation steps.

Login to generate AI explanation

CVSS v3.x Details

0.0 Low Medium High Critical 10.0
Vector String
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

References to Advisories, Solutions, and Tools

Patch Vendor Advisory Exploit Third Party Advisory

3 reference(s) from NVD

Quick Stats

CVSS v3 Score
9.8 / 10.0
EPSS (Exploit Probability)
0.1%
33th percentile
Exploitation Status
Not in CISA KEV

Weaknesses (CWE)